DPS/CPU/04/A2/2731/PT |
|
22/02/2021 |
|
22/03/2021 14:30:00 |
|
22/03/2021 14:30:00 |
|
22/03/2021 15:00:00 |
|
NA |
|
NA |
|
Qualifying Requirement : |
AS IN TENDER |
Sr. no | Item Category | Item Description | Qty | 1 | MISCELLANEOUS | Turbo Molecular/Drag pump with Controller, connection cable, mains cable, splinter shield, air cooling kit and venting valve.
Specifications for turbo molecular pump are given in the attached annexure ( against item no-1) | 2 | 2 | MISCELLANEOUS | Turbo Molecular/Drag Pump with controller connection cable, mains cable, splinter shield, air cooling kit and venting valve. Specifications for turbo molecular pump are given in the attached annexure ( against item no-2) | 2 | 3 | MISCELLANEOUS | Diaphragm Vacuum Pump as per the specifications provided in the annexure ( against item no-3) | 3 | 4 | MISCELLANEOUS | Full range vacuum gauge with two sensors (Pirani + cold cathode) in one housing. (with KF flange) as per the specifications provided in the attached annexure ( against item no-4) | 3 | 5 | MISCELLANEOUS | Full range vacuum gauge with two sensors (Pirani + cold cathode) in one housing. (with CFF flange) as per the specifications provided in the attached annexure ( against item no-5) | 3 | 6 | MISCELLANEOUS | Pirani vacuum gauge.
(Range: 1000 mbar.- 5 x 10E(-4) mbar) as as per the specifications provided in the attached annexure ( against item no-6) | 5 | 7 | MISCELLANEOUS | Dual channel controller for Pirani and full range gauge as per the specifications provided in the attached annexure( against item no-7) .
Special Notes:
• Items in S. No. 4, 5, 6 & 7 should be from the same manufacturer as they must be compatible to each other. This is treated as a part of specification.
• The suppliers/their representatives having a servicing facility in India for all items only will be considered for placement of order.
Acceptance criteria
Compatibility specification;
Items in S .No. 4, 5, 6 & 7 are inter connected each other and should be from the same manufacturer as they must be compatible to each other.
• All the Turbo molecular pumps (TMPs) need to be demonstrated to show an ultimate vacuum of at least 1x10-8 torr after two days of continuous running including baking cycles.
• The supplier must demonstrate the performance of the other items as per indent specifications at user’s site. | 3 | | | | |
| |
|
STORES OFFICER, CENTRAL STORES UNIT, DIRECTORATE OF PURCHASE AND STORES, B.A.R.C., TROMBAY, MUMBAI - 400 085. |
|