DPS/CPU/04/A3/2534 |
|
21/01/2021 |
|
24/02/2021 14:30:00 |
|
24/02/2021 14:30:00 |
|
24/02/2021 15:00:00 |
|
NA |
|
NA |
|
Qualifying Requirement : |
As per tender |
Sr. no | Item Category | Item Description | Qty | 1 | MEDICAL EQUIPMENT AND SUPPLIES | Supply, Installation & Commissioning, Training, Warranty of Ultra low temperature freezer with temperature control range of -50 degree C to -86degreeC
1. Upright Model with single door and 2 or 3 inner shelves
2. Temperature range of -50degreeC to -86degreeC in 1degreeC increment
3. Effective capacity- 500-700 liters. Vacuum insulation plus (VIP) technology for better storage capacity
4. HFC free refrigerant
5. Digital Temperature display with Control panel
6. Alarm with self-diagnosis function- High- low temperature, Power failure, Door check and Part replacement/ System failure.
7. Remote alarm contact
8. Smooth opening and closing of doors with door lock system
9. Training and installation: should be provided on site by the supplier.
10. Electric requirements: AC 230 V / 50Hz with energy efficient operation
11. Accessories: Circular Temperature recorder with chart, Inventory/ Cryoracks for samples/vials storage, 2 pairs of Cryogloves should be provided with the equipment.
12. Warranty: minimum period of 2 years
13. AMC with calibration: should be provided free for first 2 years after installation and cost of AMC with calibration thereafter should be provided at the time of quotation in the compliance statement. Temperature recorder charts should be included in AMC.
14. User list and installation: The equipment should be installed in at least few major Public/ Corporate Blood banks or laboratories or Hospitals in Mumbai and Maharashtra. User list of the same should be provided in order to take feedback from users
15. Compliance statement: including all above parameters with annexures wherever necessary is mandatory | 1 | | | | |
| |
|
ASST. STORES OFFICER, BARC HOSPITAL STORES, BARC HOSPITAL, ANUSHAKTI NAGAR, MUMBAI - 400 094. |
|